S-19-01-1214-UV101 V1.3 
  Section 5: Contraindications and Warnings   
 
5.1 Contraindications                                                             
‧ Ultrasonic Systems should not be used for 
restorative dental procedures involving the 
condensation of amalgam.   
‧ This device is designed to work with all Cavitron
®
 
(remark)  insert  with  25Khz  or  30Khz  frequency 
automatically.  For  optimum  performance  please 
use only qualified inserts supplied by Cavitron
®
 or 
Rolence
®
.   
‧ Do not use this device if the patient or operator is 
wearing a pacemaker. 
Remark: Cavitron
®
 is a registered trademark of 
Dentsply
®
 International, Inc 
 
5.2 Warnings 
‧ Persons who are fitted with cardiac pacemakers, 
defibrillators and other active implanted medical 
devices, have been cautioned that some types of 
electronic  equipment  might  interfere  with  the 
operation of the device. We recommend that the 
handpiece  and  tubing  be  kept  at  least  6  to  9 
inches (15 to 23 cm) away from any device and 
their leads during use.   
 
‧ During  using  the  unit,  make  sure  that  water  is 
flowing continuously. If the handpiece overheats 
please check the water supply, and stop using the 
unit for a while. 
 
Section 6: Infection Control 
 
6.1 General Infection Control 
Recommendations 
‧ As with all dental procedures, the use of standard 
personal  protection  equipment  (i.e.,  wearing  a 
face  mask,  eyewear,  or  face  shield,  gloves  and 
protective gown) is recommended. 
‧ For maximal operator and patient safety, carefully 
follow  section  11  system  maintenance  and  care 
information detailed  in the operating instruction 
accompanying your ultrasonic scaler. 
‧ As with high speed handpieces, and other dental 
devices, the combination of water and ultrasonic 
vibration from your ELiTEDENT
®
 Ultrasonic Scaler 
will create aerosols. With proper technique, much 
of  the  aerosol  dispersion  can  be  effectively 
controlled and minimized. Please carefully follow 
the  procedural  guidelines  in  this  manual 
regarding the use of your ultrasonic scaler.   
‧ Always flush  your  ELiTEDENT
®
 Ultrasonic  Scaler 
with highest flow before treatment. Refer to more 
information in section 10. 
‧ Clean and disinfect the handpiece sleeve between 
patients. The handpiece sleeve can be autoclaved 
up to 135℃  for at least 3 minutes. 
 
Sterilizing: 
1. Place  handpiece  sleeve  and  pouched  ultrasonic 
insert  into a steam autoclave. After  warm-up  is 
completed,  operate  at  a  sterilizing  temperature 
and pressure  of 273° F/31 psi (134°C/216 kPa) 
for  12  minutes,  followed  by  a  20-30  minute 
drying time. 
2. To maintain sterility, the insert should remain in 
the sealed pouch until it is ready for use. 
 
DO NOT USE Cold sterilization solution. 
 
6.2 Water Supply Recommendations 
‧ It  is  highly  recommended  that  all  dental  water 
supply systems should conform to applicable CDC 
(Centers for Disease Control and Prevention) and 
ADA  (American  Dental  Association)  standards, 
and  that  all  recommendations  be  followed  in 
terms of flushing, chemical flushing, and general 
infection control procedures. See sections 6.1 and 
11.