‘’’ - Followed by 3 to 5 characters.
Turbo Pump Speed
in percent of full speed.
(e.g. 1.5 = 1.5% of full speed)
(e.g. 75.8 = 75.8% of full speed)
In software revisions 3.8 to 4.0, this field is fixed at 3 characters. An
error
will occur at 100% and will only show 10%.
(‘`’ + 2 numbers).
(e.g. 25 = 25% of full speed)
In software revisions 4.1, this field is fixed at 4 characters.
(‘`’ + 3 numbers).
(e.g. 025 = 25% of full speed)
(e.g. 100 = 100% of full speed)
‘a’ - Followed by 3 to 4 characters.
Minimum Gross Testing Time (D5) in Tenths of seconds.
(e.g. 0.1 = 0.1 seconds)
(e.g. 14.2 = 14.2 seconds)
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘a’ + 4 numbers).
(e.g. 0100 = 10.0 seconds)
‘b’ -
Followed by up to 8 characters.
Total Running Time in Hours.
(e.g. 1 = 1 hour)
(e.g. 32 = 32 hours)
In software revisions 3.8 and later, this field is fixed at 7
characters.
(‘b’ + 6 numbers).
(e.g. 000047 = 47 hours)
‘c’ - Followed by 4 characters.
Minimum Roughing Time (D6) in Tenths of seconds.
This field did not exist In software revisions prior to Rev. 3.8.
(e.g. 0001 = 0.1 seconds)
(e.g. 0142 = 14.2 seconds)
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C d-xi