‘]’ - Followed by 8 characters.
Filament #, Helium Mass, Gas Type, Pressure Units, FLASH Rev.
and EPROM Rev.
Char#
1 — Filament Status (‘1’ = Filament #1, ‘2’ = Filament #2)
2 — Helium Mass (‘3’ = He3, ‘4’ = He4)
3 — Gas Type (‘0’ = Air Equiv. Units, ‘1’ = Helium Units)
4 — Pressure Units (‘0’ = mTorr, ‘1’ = mBar)
5 & 6 — FLASH Rev. (‘33’ = FLASH Rev. 3.3)
7 & 8 — EPROM Rev. (‘15’ = FLASH Rev. 1.5)
(e.g., ]14003114 = Filament On, He4, Air Equiv. Units, mTorr,
FLASH 3.1, and EPROM 1.4)
In software revisions 3.8 and later, this field is fixed at 9
characters.(‘]’ + 9 numbers. It now appends ILD auto/manual mode
status to the end).(e.g. ]140031141 = Filament On, He4, Air
Equiv. Units, mTorr, FLASH 3.1, EPROM 1.4, and ILD is in manual
mode)
‘^’ - Followed by 6 characters.
For VIC’s internal use only!
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘^’ + 4 numbers).
‘_’ - Followed by 6 characters.
For VIC’s internal use only!
In software revisions 3.8 and later, this field is fixed at 5 characters.
(‘_’ + 4 numbers).