DVM1200 VELLEMAN
- 10 -
c. Battery Replacement
Set the rotary switch to the OFF position.
Disconnect all test leads and/or any connectors from the terminals.
Open the battery compartment using an appropriate screwdriver.
Remove the battery.
Install a replacement battery (6F22).
Close the battery compartment.
6. Technical Specifications
Accuracy is specified for one year after calibration at operating temperatures of 18°C ~ 28°C with a relative humidity
at 0% ~ 75%. Accuracy specifications take the form of ± (% of rdg + number of least significant digits).
Environmental Conditions 1000V CAT. II and 600V CAT. III
Pollution Degree 2
Altitude < 2000m
Operating Temperature 0°C~40°C or 32°F~122°F (< 80% RH, < 10°C)
Storage Temperature -10°C~60°C or 14°F~140°F (< 70% RH, battery removed)
Temperature Coefficient 0.1x / C° (< 18°C or > 28°C)
Max. Voltage between Terminals and Earth 700VAC RMS or 1000VDC
Fuse Protection μA and mA, F750mA / 250V, 5 x 20mm
A, F10A / 600V, 6.3 x 32mm
Sample Rate 3x/sec for digital data
Display 3
5/6
digits LCD with automatic indication of functions and symbols
Over Range Indication yes (“OL”)
Low Battery Indication yes ( )
Polarity Indication -displayed automatically
Data Hold yes
Backlight white LEDs
Auto Power-Off yes
Power Supply 9V battery
Dimensions 180 x 85 x 40mm
Weight ± 310g (with battery)
Accessories 1 x K-type thermocouple, 2 x test leads, 1 x optical USB cable, PC-link
software, 1 x 9V battery, manual
Voltage
DC Voltage
Range
Resolution
Accuracy
600mV
0.1mV
± (0.5% of rdg ± 5 digits)
6V
1mV
± (0.8% of rdg ± 5 digits)
60V
10mV
600V
100mV
1000V
1V
± (1.0% of rdg ± 10 digits)
To avoid false readings, which could lead to possible electric shocks or personal injury,
replace the battery as soon as the battery indicator appears. Before battery replacement,
disconnect all test leads and/or ant connectors from any circuit under test.