Display Error Codes
VELOPEX INTRA-X 17
User Guide
Display
Characters
Description Action
E03 Dryer temperature sensor s/c Call for Service
E04
Dryer temperature sensor o/c
or connection to Main PCB
Call for Service
E05 Fixer temperature sensor s/c Call for Service
E06 Fixer temperature sensor o/c Call for Service
E07
Developer temperature sensor
s/c
Call for Service
E08
Developer temperature sensor
o/c
Call for Service
E10 Fan Sensor Inactive Call for Service
E11 Drive turns count not zero Call for Service
E12 Drive motor overloaded
1. Switch off the Machine .
2. Clean the Gears on the
Modules.
3. Switch on the Machine.
4. If still E12 - Call for Service.
E13
Transport motor Sensor
Inactive
or Modules seized
1. Switch off the Machine
2. Clean the gears and belts
on the Modules.
3. Switch on the machine
4. If still E13 - Call for Service
E14
Excessive Fixer / Developer
temperature
Call for Service
E15 Dryer temperature too high Call for Service
E18 Dryer power duty cycle too low Call for Service
E33
Watchdog relay test failure
(Low)
Call for Service
E34
Watchdog relay test failure
(High)
Call for Service
tLO
Either Developer or Fixer tank
temperature is low
Wait for the correct
temperature
LId Lid switch open Close lid