User Guide
VentMed
interference in nearby electronic equipment.
RF emissions
CISPR11
Class B
This device is suitable for use in all
establishments, including domestic
Harmonic emissions Class A
IEC61000-3-2
establishments and those directly connected to
the public low-voltage power supply network
Voltage fluctuation Conformity that supplies buildings used for domestic
/flicker emissions
IEC61000-3-3
purposes.
Guidance and manufacturer's declaration - electromagnetic immunity this
device is intended for use in the electromagnetic environment specified below. The
user of the device should make sure that it is used in such an environment.
Immunity test IEC60601
level
test Compliance level
Electromagnetic
environment-guidance
Floor should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic
discharge
(ESD)
±6KV contact
±8KV air
±6KV contact
±8KV air
IEC61000-4-
2
Electrical fast ±2KV for
±2KV for
Mains power quality
transient /
burst
power supply lines power supply lines should be that of a typical
±1KV for
±1KV for
home or hospital.
IEC61000-4- input/output lines
input/output lines
33