CENTRA-MDS/LDS/RDS - US
Operator Manual

Questions and Answers

  • M
    Mary BoyleAug 20, 2025
    What to do if there is an alarm and flashing quality value on Veolia ELGA CENTRA-RDS-US?
    • T
      Timothy CarterAug 20, 2025
      If you observe an alarm and flashing quality value on your Veolia Water Filtration System, first mute the alarm. Verify that the alarm set value is correct. If your system has an ion-exchange/purification media cylinder, ensure it is properly in line and not isolated from the distribution loop. If necessary, replace the ion-exchange cylinder. If the problem continues, further assistance may be needed.
  • R
    Ryan Lee Jr.Aug 22, 2025
    What does '=== M?.cm (RDS/RDS(HF))' mean on Veolia Water Filtration Systems?
    • V
      Veronica MannAug 22, 2025
      If your Veolia Water Filtration System displays '=== M?.cm (RDS/RDS(HF))', it indicates that a feature is out of the measurement range. Allow the unit to recirculate. If the purification technology in the recirculation loop is insufficient, verify the design with Technical Support. Also, check that the quality sensor is connected. If fitted, replace the ion-exchange/purification media cylinder. If the problem persists, further assistance may be needed.
  • M
    Michael RobertsAug 25, 2025
    What to do if there is a high water temperature alarm on Veolia ELGA CENTRA-RDS-US?
    • D
      Dr. Lisa Reid MDAug 25, 2025
      If you receive a high water temperature alarm on your Veolia Water Filtration System, first check that the correct alarm point is set. Next, check if the feedwater temperature has risen suddenly. Dispense some water from the distribution loop to allow cold water to be drawn into the unit, and then check the feedwater temperature again. If the problem continues, further assistance may be needed.
  • T
    Thomas FisherAug 26, 2025
    How to clear the 0.2µm filter change reminder alarm on Veolia Water Filtration Systems?
    • Z
      zhardingAug 26, 2025
      If your Veolia Water Filtration System displays a 0.2µm filter change reminder alarm, mute the alarm and replace the filter. If the problem persists, further assistance may be needed.
  • V
    Victoria JonesAug 28, 2025
    How to clear the UV change reminder on Veolia ELGA CENTRA-RDS-US?
    • C
      Charles TerryAug 28, 2025
      If your Veolia Water Filtration System displays a UV change reminder, mute the alarm and replace the UV lamp. If the problem persists, further assistance may be needed.
  • T
    turnerkellyAug 29, 2025
    How to clear the sanitization reminder alarm on Veolia ELGA CENTRA-RDS-US Water Filtration Systems?
    • M
      Mrs. Cynthia Sanders MDAug 30, 2025
      If your Veolia Water Filtration System displays a sanitization reminder alarm, mute the alarm and initiate the Sanitization Procedure. If the problem persists, further assistance may be needed.
  • D
    Darlene ParkSep 1, 2025
    What to do if the reservoir low level alarm is on for Veolia ELGA CENTRA-RDS-US Water Filtration Systems?
    • T
      Tara HernandezSep 1, 2025
      If your Veolia Water Filtration System shows a reservoir low level alarm, check that the feedwater supply is connected and ensure the demand matches the water production. If the problem persists, further assistance may be needed.
  • N
    Nicholas VanceSep 3, 2025
    What to do if the reservoir level disconnect fault alarm is on for Veolia ELGA CENTRA-RDS-US Water Filtration Systems?
    • A
      Amanda TranSep 3, 2025
      If your Veolia Water Filtration System displays a reservoir level disconnect fault alarm, mute the alarm and ensure the control lead from the reservoir is properly connected. If the problem persists, further assistance may be needed.
  • J
    John RamirezSep 4, 2025
    Why is there reduced flow from the distribution loop in Veolia Water Filtration Systems?
    • B
      Brittany OrtizSep 5, 2025
      If your Veolia Water Filtration System experiences reduced flow from the distribution loop, check the pressure drop across the 0.2µm filter. Replace the filter if the pressure drop is higher than 0.5bar. If this doesn't solve the issue, the pump may be worn.
  • K
    Katherine JacksonSep 7, 2025
    How to reduce the noise level of Veolia Water Filtration Systems?
    • J
      John AdamsSep 7, 2025
      If the noise level of your Veolia Water Filtration System is greater than specified, open the front door and secure the pipework to stop vibration. Also, check that air is not being drawn into the recirculation loop. If the problem continues, further assistance may be needed.

Summary

Health and Safety Notes for CENTRA

Electrical Safety Precautions

Essential safety precautions related to the electrical supply, isolation, and earthing of the CENTRA unit.

Pressure Safety Guidelines

Guidelines for safely isolating and releasing pressure before maintenance or work on the unit.

Product and Process Description

CENTRA Process Description

Explains the water flow and component functions within the CENTRA LDS, RDS, and HFV/HFR systems.

CENTRA Technical Specifications

Details the feedwater, drain, dimensions, weights, connections, and product water specifications.

Installation Instructions for CENTRA

Positioning the CENTRA Unit

Guidelines for selecting a suitable environment and location for installing the CENTRA unit.

Connecting the CENTRA Hydraulics

Details the hydraulic connections required for different CENTRA models.

CENTRA Controls and Operation

CENTRA PASSkeys System

Explains the different types of PASSkeys and their functions for system access and control.

CENTRA System Preferences Setup

Guidance on setting up system preferences like clock, date, and alarms via the control panel.

Setting Up Replacement Timers; Reminders

Procedure to enter the consumable replacement timer setup for various filters and UV lamps.

CENTRA Commissioning Procedures

CENTRA Operation Modes

CENTRA Maintenance Procedures

CENTRA Sanitization Procedures

Trouble Shooting CENTRA Issues

Key to CENTRA Control Panel

CENTRA Alarm Conditions

Details of potential faults, their causes, and actions for common alarms displayed by the CENTRA.

CENTRA Replacement Timers

Information on setting reminders for filter, UV lamp, and sanitization replacements.

Veolia ELGA CENTRA-RDS-US Specifications