DIGITAL MULTIMETER VC9808+
OPERATION MANUAL
1. GENERAL
The instrument is a stable and good performance digital multi-meter driven by
battery. It uses the LCD with 26mm-high figure to make the reading clear. Unit
symbol displaying, and the function of backlight and overload protection make
operation is convenient.
The instrument has the function of measuring DCV, ACV, DCA, ACA, resistance,
capacitance, inductance, temperature and frequency, and diode, triode and
continuity test. The instrument takes dual-integral A/D converter as key point,
is an excellent tool.
2. SAFETY NOTES
This series meter meets the standard of IEC1010. Please read below notes
carefully before operation.
1Do not input a limited voltage which RMS is over DC 1000V or AC 750V when
measuring voltage.
2Voltage less than 36V is a safety voltage. When measuring voltage higher
than DC 36V, AC 25V, check the connection and insulation of test leads to avoid
electric shock.
3Be sure to keep the test leads off the testing points when converting function
and range
4Select correct function and range to avoid fault operation.
5When measuring current, do not input current over 20A.
6Safety symbols
” exists high voltage, ”GND, ” dual insulation,
” must refer to manual, ”low battery indication.
3. SPECIFICATION
1) GENERAL
1-1Displaying : LCD displaying
1-2Max. indication: 19993 1/2, auto polarity indication
1-3Measuring method: dual slope A/D transfer
1-4Sampling rate: approx. 3 times/sec
1-5Over range indication: MSD displays “OL” or -OL”
1-6.Low battery indication: “ ” symbol displays
1-7Operation: 0~40℃,relative humidity <80%
1-8Storage: -10~50℃,relative humidity <80%
1-9Powerone 9V battery (6F22 or equivalent
1-10Size: 185×93×35 mm
1-11Weight: approx.290g (including a 9V battery)
1-12. Accessories: Operation Manual, holster, gift box, test leads, a pair of
TP01 thermocouple, 9V battery.
2) TECHNICAL DATA
2-1.Accuracy: ±(a% × reading + dgts at 23 ± 5℃, relative humidity<75% .
One year guarantee since production date.
2-2. TECHNICAL DATA
2-2-1. DC Voltage
Range
Accuracy
Resolution
200mV
±(0.5%+3)
0.1mV
2V
1mV
20V
10mV
200V
100mV
1000V
±(1.0%+5)
1V
Input resistance:10MΩ for all ranges
Overload protectionRange 200mV: DC 250V or AC peak value
Other ranges: DC 1000V or AC peak value
2-2-2. AC Voltage
Range
Accuracy
Resolution
200mV
±(0.8%+5)
0.1mV
2V
1mV
20V
10mV
200V
100mV
750V
±(1.2%+5)
1V
Input resistanceIt is 1MΩ at range 200mV or 2V, and be 10MΩ at others
Overload protection: DC 250V or AC peak value at range 200mV
DC 1000V or AC peak value at others
Frequency responseRange less than 200V: 40~400Hz
Range 750V: 40~100Hz
Displaying: RMS of sine wave (mean value response)
2-2-3. DCA
Range
Accuracy
2mA
±(0.8%+3)
20mA
200mA
±(1.2%+4)
20A
±(2.0%+5)
Max. measuring voltage drop: 200mV
Max. input current: 20A (within 10 seconds)
Overload protection: 0.2A / 250V fast-blown fuse, range 20A infused.
2-2-4. ACA
Range
Accuracy
Resolution
2mA
±(1.0%+5)
1uA
20mA
±(2.0%+5)
10uA
200mA
100uA
20A
±(3.0%+10)
10mA
Max. measuring voltage drop: 200mV
Max. input current: 20A (within 10 second
Overload protection 0.2A / 250V quick-action fused, Range 20A infused.
Frequency response40~200Hz
Displaying: RMS of sine wave (mean value response)
2-2-5. Resistance(Ω)
Range
Accuracy
Resolution
200Ω
± (0.8%+5)
0.1Ω
2kΩ
± (0.8%+3)
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
20MΩ
± (1.0%+15)
10kΩ
2000
± [ 5% ( reading 10 )+
20 ]
1MΩ
Open voltageless than 3V
Overload protectionDC 250V or AC peak value
NOTE:
AAt range 200 Ω, short-circuit the test leads to measure the wire resistance,
then, subtracts it from the real measurement.
BAt range 2000 MΩ, when the test leads short circuit, LCD displays 10,
It’s normal and doesnt affect the accuracy. Please deduct the value from
the real measurement. For example: the tested resistance is 1000 , the
reading should be 1010 MΩ, and the correct value is reduce 10 from the LCD
reading value, that is 1010-10=1000MΩ;
CThe reading be displayed slowly when the measurement is more than 1 .
Please wait it to be stable.
2-2-6. Capacitance
Range
Accuracy
Resolution
20nF
±(2.5%+20)
10pF
200nF
100pF
2uF
1nF
20uF
10nF
2000uF
±(5.0%+5)
1uF
Test frequency: 100Hz
Overload protectionDC 36V or AC peak value
2-2-7. Inductance (L)
Range
Accuracy
Resolution
2mH
±( 2.5%+20)
1uH
20mH
10uH
200mH
100uH
2H
1mH
20H
10mH
Test frequency100Hz
Overload protectionDC 36V or AC peak value
2-2-8. Temperature
Range
Accuracy
Resolution
(-40 ~ 1000)
±(0.8%+4) < 400℃
±(1.5%+15) ≥ 400℃
1℃
K-type thermocouple ( banana shape plug )
2-2-9. Frequency
Range
Accuracy
Resolution
2kHz
±(0.5%+4)
1Hz
20kHz
10Hz
200kHz
100Hz
2000kHz
1kHz
10MHz
10kHz
Input sensitivity1V RMS
Overload protectionDC 250V or AC peak value (within 10 seconds)
2-2-10. Diode and continuity testing
Range
Reading
Condition
Forward voltage drop of diode
Forward DCA is approx.
1mA, the backward
voltage is approx 3V
Buzzer makes a long sound while
resistance is less than 3±1
Open voltage is approx.
3V
Overload protectionDC 250V or AC peak value
CAUTIONDO NOT INPUT VOLTAGE AT THIS RANGE!
2-2-11. Transistor hFE DATA TEST
Range
Displaying range
Test condition
hFE NPN or PNP
0 ~ 1000
Basic current is approx.
10uA Vce is about 3V
4. OPERATION
4-1Panel description
1. LCD: display the measuring value and unit.
2-1. POWER switch: turn on/off the power.