OPERATION MANUAL
The meter is a high performance, high accuracy, 3 3/4-digit, 32mm digit high by
LCD displaying meter. The function includes DCV, ACV, DCA, ACA,
resistance ,capacitance, frequency, temperature,duty cycle measurement, triode,
diode and continuity test. Also with the function of unit signal displaying ,relative
value measuring ,auto range/ manual range switch ,auto power off and warning
function etc.
SAFETY NOTES
The meter meets with IEC1010standard. Read the notes before operating.
①Before measuring, check the test leads carefully.
②Do not input a value over allowable limited when measuring to avoid electric
shock meter damage.
③Be careful when measuring voltage over DCV60V, ACV40V.
④To select correct function.
⑤The test leads should be taken away from the testing point when changing
function.
⑥Never take measurement when the test leads are in current terminal.
⑦Do not try to modify the circuit.
⑧Safety symbols
“ “ dangerous voltage
“ “ GND
“ “ dual insulation
“ “ refer to the operation manual
“ “ low battery
SPECIFICATION
1. General
1-1,Display: 3 3/4digit LCD with a max readings of 3999, Automatic
negative polarity indication.,
1-2.Measuring method: Dual-slop integrating A/D converter system.
1-3.Sampling rate: Approx. 3times/second.
1-4 Over range indication: “ OL” display.
1-5.Low battery: The “ ”is display.
1-6.Operating environment : Temperature (0℃ to 40℃), humidity<80%RH.
1-7.Storage environment: Temperature (-10℃ to 50℃), humidity<80%RH.
1-8.Power: 2pcs 1.5V AAA 7# battery.
1-9.Dimension: 185(L)X93(W)X35(H)mm.
1-10Weight: Approx. 290g(including battery).
1-11.Accessories: manual, gift box ,holster ,TP01 temperature probe, test leads and
2pcs 1.5V battery.
2. Technical
2.1 Accuracy: ±(a% × reading + digits) at 23±5℃,<75%RH.
2.2 DCV
Input impedance: 400mv range>40MΩ,other range is 10MΩ.
Overload protection: 1000V DC or 700V AC peak value
2.3 ACV
Input impedance: 400mv range>40MΩ,other range is 10MΩ.
Overload protection: 1000V DC or 700V AC peak value
Frequency response: 750V range: 40~100Hz, other range: 40~400Hz
Displaying: sinewave RMS(average value response)
2.4 DCA
Max. measuring volt drop: 400mV under mA range, 200mV under A
range.
Max. input current: 20A(less than 15sec.)
Overload protection: 0.5A/250V fuse, 13A/250V fuse.
2.5 ACA
Max. measuring volt drop: 400mV under mA range, 200mV under A
range.
Max. input current: 20A(less than 15sec.)
Overload protection: 0.5A/250V fuse, 13A/250V fuse.
Frequency response: 40~100Hz under 10A range, 40~400Hz at other
range.
2.6 RESISTANCE