12
Recommended separation distances between portable and mobile
RF communications equipment and the [ME EQUIPMENT or
ME SYSTEM]
The Infusion is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Infusion
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
Infusion as recommended below, according to the maximum output power of the
communications equipment
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
N/A
Table 7 – Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY
– for LIFE-SUPPORTING ME EQUIPMENT and ME SYSTEMS that are specified for use only in a
shielded location
N/A
Table 8 – Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY
– for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING and are specified for
use only in a shielded location.
N/A because the device is not specified for use only in a shielded location.