Table 4 – Guidance and Manufacture’s declaration – electromagnetic IMMUNITY – for ME
EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Table 6 – Recommended separation distances between portable and mobile RF
communications equipment and the ME EQUIPMENT or ME SYSTEM – for ME
EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
N/A
Table 5 – Recommended separation distances between portable and mobile RF
communications equipment and the ME EQUIPMENT or ME SYSTEM – for LIFE-SUPPORTING
ME EQUIPMENT and ME SYSTEMS
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Infusion is used
exceeds the applicable RF compliance level above, the Infusion should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Infusion. b Over the frequency range 150
kHz to 80 MHz, field strengths should be less than [V1] V/m.