Table 4 – Guidance and Manufacture’s declaration – electromagnetic IMMUNITY – for ME
EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
N/A
Table 3 – Guidance and Manufacture’s declaration – electromagnetic IMMUNITY – for
ME EQUIPMENT and ME SYSTEMS that are LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The Infusion is intended for use in the electromagnetic environment specified below.
The customer or the user of the Infusion should assure that it is used in such an
environment.
IMMUNITY test IEC 60601 TEST
LEVEL
Compliance
level
Electromagnetic
environment – guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC 61000-
4-3
3 Vrms 150 kHz to 80
MHz
3 V/m 80 MHz to 2,5
GHz
[3] V
[3] V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Infusion, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as deter-mined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol: