Guidance and manufacturer’s declaration – electromagnetic immunity
The Infusion is intended for use in the electromagnetic environment specified below.
The customer or the user of the Infusion should assure that it is used in such an
environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air ± 6 kV contact ± 8
kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power supply
lines ± 1 kV for input/
output lines
± 2 kV for power
supply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5 ± 1 kV line(s) to line(s) ±
2 kV line(s) to earth
± 1 kV line(s) to
line(s) ± 2 kV line(s)
to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
<5 % UT
(>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in
UT) for 5 s
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Infusion
requires continued
operation during power
mains interruptions, it is
recommended that the
Infusion be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
3 A/m 3 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.