5. Service Mode
52 EzRay Air Portable (Model: VEX-P300) User Manual
5.3.2 Exposure Time Settings (for each patient and tooth type)
Select a service mode from n.02 to n.09 to set up an exposure time for each patient
and tooth type. See the table below for more information.
Service Mode No. Item
n.02 Adult Incisor
n.03 Adult Canine
n.04 Adult Molar/ Premolar
n.05 Adult Bitewing
n.06 Child Incisor
n.07 Child Canine
n.08 Child Molar/ Premolar
n.09 Child Bitewing
To set the exposure time for a specific patient and tooth type as described above,
1. Select the Service Mode number according to the patient and tooth type (i.e.,
n.06 for Child Incisor)
2. After entering the mode, turn the jog dial to adjust the exposure time.
3. Press the jog dial again and hold it until you hear a beep.
4. Restart the device to check if the new setting is saved.