5. Service Mode
EzRay Air Portable (Model: VEX-P300) User Manual 53
ENGLISH
5.3.3 User Default Settings (for each patient and tooth type)
User default settings are activated as the device starts. See the table to change
each service mode number aligned with the patient or tooth type.
Service Mode No. Item
n.10 Adult
n.11 Child
n.12 Incisor
n.13 Canine
n.14 Molar/Premolar
n.15 Bitewing
To set the default setting for a specific patient type or a tooth type,
1. Select a service mode number between n.10 (Adult) and n.11 (Child) for a
specific patient type and from n.12 to n.15 for a tooth type. For instance, when
select n.10 by pressing the jog dial to set the patient type ‘adult’ as default, then
“YES” is displayed on the Control panel.
2. Press the jog dial again and hold it until you hear a beep.
3. Restart the device to check if the new setting is saved correctly.