Settings MU 6100 H
62 ba77077e02 09/2014
10.2 Sensor-independent settings
10.2.1 System
To open the Storage & config menu, press the <OK
_> key in the measured
value display. After completing the settings, switch to the measured value dis-
play with <MODE>.
Default settings are printed in bold.
Menu item Possible
setting
Explanation
System / General /
Language
Deutsch
English
(more)
Selects the menu language
System / General /
Beep
On
Off
Switches on/off the beep on
keystroke
System / General / Illu-
mination
Auto
On
Off
Switches the display illumina-
tion on/off
System / General /
Contrast
0 ... 50 ...
100 %
Changes the display contrast
System / General /
Switchoff time
10 min ... 1h
... 24 h
Adjusts the shut-off time
System / General /
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are dis-
played with the selected unit.
System / General /
Stability control
On
Off
Switches on or off the auto-
matic stability control during
measurement
System / Interface /
Baud rate
1200, 2400,
4800, 9600,
19200
Baud rate of the data interface
System / Interface /
Output format
ASCII
CSV
Output format for data trans-
mission (see section 12 T
RANS-
MITTING DATA (USB INTERFACE),
page 72)
System / Interface /
Decimal separator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System / Interface /
Output header
Output of a header for Output
format: CSV