What to do if... MU 6100 H
80 ba77077e02 09/2014
No stable measured
value
Obviously incorrect
measured values
14.2 Dissolved oxygen
Error message
OFL, UFL
The measured value is outside the measuring range.
Cause Remedy
Electrode:
Junction contaminated Clean the junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Temper if necessary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high Use a suitable electrode
Organic liquids Use a suitable electrode
Cause Remedy
Electrode:
Electrode unsuitable Use a suitable electrode
Temperature difference between
buffer and test sample too great
Adjust temperature of buffer or
sample solutions
Measurement procedure not suit-
able
Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
The measured value is outside
the measuring range of the meter
Use a suitable D.O. sensor