16
• Door switch or sensor error
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
Turn the power switch off and back on to reset.
* If problem cannot be resolved, call VWR Service Department
• Door switch or sensor error
• Door latch is not operating properly
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
• Door switch or sensor error
• Door latch is not operating properly
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
Instrument is vibrating
excessively or making
unusual noise
Perform the recommended corrective action as listed in
E8.
• Mechanical failure or damage
Turn the power off immediately and call VWR Service
Department.
* If problem cannot be resolved, call VWR Service Department.