16
E4
• Door switch or sensor error
Corrective
Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
E5
• Temporary system error
Corrective
Action
Turn the power switch off and back on to reset.
* If problem cannot be resolved, call VWR Service Department
E6
• Door switch or sensor error
Door does not open
• Door latch is not operating properly
E7
Corrective
Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
• Door switch or sensor error
Door does not close
• Door latch is not operating properly
E8
Corrective
Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency
Door Open function to open the door manually to save
samples. Call VWR Service Department.
E9
• Rotor is not balanced
Instrument is vibrating
excessively or making
unusual noise
Corrective
Action
Perform the recommended corrective action as listed in
E8.
• Mechanical failure or damage
Corrective
Action
Turn the power off immediately and call VWR Service
Department.
* If problem cannot be resolved, call VWR Service Department.