EasyManua.ls Logo

Watts Benchmark Aerco OMM-0122 - Page 64

Watts Benchmark Aerco OMM-0122
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Benchmark 750-3000 Boiler Installation & Startup Manual
SECTION 4 INITIAL START-UP
OMM-0121_D GF-205 5/9/2019 Technical Support (800) 526-0288 Mon-Fri, 8 am - 5 pm EST Page 64 of 122
NATURAL GAS Combustion Calibration Instructions
15. On Benchmark 3000 units only, once the manifold gas pressure is within the range in
table 4-1, record this value; it will be used later, in Section 5.2.2: Low Pressure Gas Test,
and Section 5.3.2: High Pressure Gas Test.
16. Press the Down▼ arrow key until SET Valve Position appears on the display, then press
CHANGE.
17. Press the Up ▲ arrow key until the SET Valve Position reads 100%, then press ENTER.
18. With the valve position at 100%, insert the combustion analyzer probe into the exhaust
manifold probe opening (see Figure 4-2a or 4-2b in Section 4.2.3) and allow enough time
for the combustion analyzer reading to stabilize.
19. Compare the combustion analyzer’s oxygen readings to the O
2
Sensor value displayed in
the Operating menu on the C-More Controller. The offset value in the Calibration menu can
be adjusted by ±3% to make the on-board O
2
sensor match the value from the combustion
analyzer. If the values differ by more than ±3% and your combustion analyzer is correctly
calibrated, the on-board O
2
sensor may be defective and need to be replaced.
20. Compare the measured oxygen level to the oxygen range shown in Table 4-2. Also, ensure
that the nitrogen oxide (NO
x
) and carbon monoxide (CO) readings do not exceed the
values shown in Table 4-2. If your target NO
x
value is 9 ppm or less, use the values in the
Ultra-Low NO
x
columns. If you are not in a “NO
x
-limited” area and/or do not have a NO
x
measurement in your analyzer, set the oxygen (O
2
) at 5.5% ± 0.5%.
TABLE 4-2: NATURAL GAS Calibration Readings 100% Valve Position
Model
Standard NO
x
Ultra-Low NO
x
Carbon
Monoxide (CO)
Oxygen (O
2
)
%
Nitrogen
Oxide (NO
x
)
Oxygen (O
2
)
%
Nitrogen
Oxide (NO
x
)
750 5.5% ± 0.2% 20 ppm 6.0% ± 1.0% ≤9 ppm <100 ppm
1000
5.5% ± 0.2%
≤20 ppm
6.0% ± 1.0%
≤9 ppm
<100 ppm
1500 5.2% ± 0.2% ≤20 ppm 5.7% ± 1.0% ≤9 ppm <100 ppm
2000 6.0% ± 0.2% ≤20 ppm 6.0% ± 1.0% ≤9 ppm <100 ppm
2500 5.6% ± 0.2% ≤20 ppm - - <100 ppm
3000 5.1% ± 0.2% ≤20 ppm - - <100 ppm
3000 DF 5.3% ± 0.2% ≤20 ppm - - <100 ppm
21. Next, press the Down ▼ arrow key until CAL Voltage 100% is displayed.
22. Press the CHANGE key and observe that CAL Voltage 100% is flashing.
23. The oxygen level at the 100% valve position should match the value in Table 4-2. Also,
ensure that the NO
x
and CO readings do not exceed the values in Table 4-2.
24. If the oxygen level is not within the specified range, adjust the level using the and
arrow keys. This will adjust the output voltage to the blower motor as indicated on the
display. Pressing the Up arrow key increases the oxygen level and pressing the down
▼ arrow key decreases the oxygen level.
25. Once the oxygen level is within the specified range at 100%, press the ENTER key to store
the selected blower output voltage for the 100% valve position. Record all readings on the
Combustion Calibration Sheets provided.

Table of Contents

Related product manuals