4. Once the airway has been secured,
switch to IPPV ventilation via the menu
button and “Ventilation mode”.
5. Ventilate continuously in “IPPV”
mode. The device shows the parameters
for IPPV ventilation.
Info: If the set pMax is reached during
CPR, MEDUVENT Standard restricts
inspiratory pressure.
Therapydisruptedbyexcessivelyhighow!Iftheowexceedsthemaximumper-
mitted value of 15 l/min, the pressure relief valve may open unintentionally during
inspiration, putting therapy at risk.
This might injure the patient.
Feedinoxygenonlyatamaximumowof15l/min.