Preparing for the function check
1. Check battery status: The rechargeable battery must be fully charged.
If necessary: Charge rechargeable battery.
2. Check the following parts for external damage:
- Device
- Labels on the device
- Connectors and cables
- Breathing circuit
- Accessories
3. If necessary: Replace damaged parts.
4. Check the patient valve of the reusable breathing circuit.
5. If necessary: Replace the breathing circuit.
6. Checkthellleveloftheoxygencylinder.
7. If necessary: Replace the oxygen cylinder.
Performing a function check
You can perform the function check with the following test lungs:
Testing bag WM 1453
Testing bag WM 1454
EasyLung for WEINMANN Emergency WM 28625
Requirement
The device is disconnected from the patient.
The rechargeable battery is fully charged.
There is an SD card in the SD card slot.
• Thehygienelterisinsertedintheltercompartment.
The oxygen supply is shut off.
The function check has been prepared.
The device is switched on.