WM 66581a 11/2017
OXYWAY EN 21
5 Hygienic preparation
5 Hygienic preparation
5.1 General instructions
This product may contain disposable items. Disposable
items are intended to be used only once. So use these
items only once and do not reprocess them. Reprocessing
disposable items may impair the functionality and safety of the
product and lead to unforeseeable reactions as a result of
ageing, embrittlement, wear, thermal load, the effects of
chemical processes, etc.
Wear suitable protective equipment for hygienic preparation
work.
You can find further information about hygienic preparation
and a list of all the suitable cleaning agents and disinfectants in
a brochure on the Internet at
www.weinmann-emergency.com.
Please refer to the instructions for use supplied with the
disinfectant used.
Always carry out a function check after the hygienic
preparation (see “6.2 Performing a function check”, page 23).
5.2 Intervals
Part concerned Interval Maintenance by
Pressure reducer
Cleaning as required (e.g., after
use in dusty environment)
User/operator
Disinfection before every new
patient