8 Maintenance
8.5 Replacing the sealing ring
Requirement The pressure reducer is removed (see “4.2 Removing the
pressure reducer”, page 16).
Wash your hands thoroughly in advance.
The small screwdriver must be clean, dry and free from any oil
or grease.
1. Lift the sealing ring gently using the screwdriver.
2. If necessary: Clean the groove with a lint-free, dry or moistened
cloth.
3. If necessary: Remove any residual moisture with the lint-free
cloth.
4. Press the new sealing ring into the groove of the hand-screwed
connection gently and without using any tools.
Result The sealing ring is replaced.