Directions for Use Chapter 7 Reference 81
Nurse Call
Switch current 1A maximum
Switch voltage 30V ac/dc maximum
Isolation 1500 Vrms
Alarm relay Energized during patient alarm
Cable Welch Allyn part number 008-0634-XX, customized to connect to the local Nurse Call system.
Cable properties:
10 feet (3 meters) long.
Not shielded.
One end not terminated (to be customized on site).
One end terminated with a connector which mates to the Nurse Call connector (shown below)
on the monitor.
4 (not connected)
1 (Black)
Normally Open
3 (Green)
Normally Closed
2 (Red) Arm