48 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph
Recording a Stat Auto ECG
A stat Auto ECG is an immediate printout.
Stat mode bypasses patient data entry, as shown in Figure 31 on page 48 in Directions for
Use. A temporary ID number is assigned to the patient to identify stat tests.
To Record a Stat Auto ECG
1. Press and hold the Auto ECG key .
The electrocardiograph begins acquiring ECG data. After it has acquired 10 seconds of
quality data, it prints a report.
2. Go to Step 4 on page 50.
WARNING To avoid the risk of associating ECG reports with the wrong patients,
each report must identify the patient. If any ECG report does not identify the
patient, write the patient identification information on the report immediately
following the ECG test.
Note
Write the patient identification information on the printed report immediately after
the ECG test to avoid associating the ECG report with the wrong patient.
ECG report Print Copy feature will not be available if no Patient Name or ID are
entered.