1.10
POWER AND POWER LINE
POWER RATING
_-
9 Watts power module operated while simultaneously charging battery.
0.6 Watts battery power
OPERATING MODES
Power Line (Mains)
Battery
-
Rechargeable
(NiCad)
or Non-Rechargeable (Alkaline)
LINE (MAINS) VOLTAGE
115 V or 220 V
LINE (MAINS) VOLTAGE VARIATION
+lo%
LINE (MAINS) FREQUENCY RANGE
50-60 Hz
LINE (MAINS) FREQUENCY VARIATION
+5%
1.11
BATTERY PACK VOLTAGE
BATTERY VOLTAGE OPERATING RANGE
7.0 v to 9.0 v
BATTERY PACK TYPICAL OPERATING LIFE
NiCad:
22 Hours
Alkaline:
45 Hours
1.12 ENVIRONMENTAL CONDITIONS
OPERATING TEMPERATURE RANGE
15 to 40 Degrees Celsius
SHELF TEMPERATURE RANGE
Power Module Operated:
-40 to 60 Degrees Celsius
Battery Operated: -40 to 40 Degrees Celsius
RELATIVE HUMIDITY RANGE
5% to 90%