Error Code
--
E26
E27
E28
&
E29
E31
E32
E33
E34
E35
E36
E37
E38
E39
Probable Cause and Solution
EEPROM CHECKSUM ERROR
-
Data in the Mirror Memory
is in error. If not recoverable perform
calibration. If the problem is not resolved by
calibration replace
U503.
EPROM TO EEPROM DATA LOAD ERROR
-
If not
recoverable replace
U503
then
U505.
CALIBRATION TYPE READ ERROR
-
Should occur only
upon entry to Calibration Mode.
If not recoverable
the problem may be related to
S2,
S6,
U509,
U508,
U506,
or
U503.
BATTERY SHUT DOWN LEVEL READ ERROR
-
If not
recoverable perform Battery Low/Shut Down
Calibration (Calibration Section 4.9.4).
If the
problem persists replace
U503.
BATTERY WARNING LEVEL READ ERROR
-
If not
recoverable perform Battery Low/Shut Down
Calibration (Calibration Section 4.9.4). If
problem persists replace
U503.
STACK OVERFLOW ERROR
-
Should always be
recoverable.
If the error persists replace
U503
or
U505.
BAD POINTER TO EEPROM ROUTINES
-
If not recoverable
replace
U505
then
U503.
INVALID KEY CODE ERROR
-
If not recoverable replace
U505
then
U503.
INVALID KEY SOURCE ERROR
-
If not recoverable
replace
U505
then
U503.
INVALID QUEUE POINTER ERROR
-
If not recoverable
replace
U505
then
U503.
BATTERY VALUE OUT OF RANGE
-
If not recoverable
perform Battery Low/Shut Down Calibration
(Calibration Section 4.9.4).
If problem persists
replace
U503.
ERROR PERTAINS TO FACTORY CALIBRATION PORT
-
If not
recoverable replace
U503.