EasyManua.ls Logo

Welch Allyn Grason-Stadler GSI 17 Series - Troubleshooting Error Codes E26-E39

Welch Allyn Grason-Stadler GSI 17 Series
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Code
--
E26
E27
E28
&
E29
E31
E32
E33
E34
E35
E36
E37
E38
E39
Probable Cause and Solution
EEPROM CHECKSUM ERROR
-
Data in the Mirror Memory
is in error. If not recoverable perform
calibration. If the problem is not resolved by
calibration replace
U503.
EPROM TO EEPROM DATA LOAD ERROR
-
If not
recoverable replace
U503
then
U505.
CALIBRATION TYPE READ ERROR
-
Should occur only
upon entry to Calibration Mode.
If not recoverable
the problem may be related to
S2,
S6,
U509,
U508,
U506,
or
U503.
BATTERY SHUT DOWN LEVEL READ ERROR
-
If not
recoverable perform Battery Low/Shut Down
Calibration (Calibration Section 4.9.4).
If the
problem persists replace
U503.
BATTERY WARNING LEVEL READ ERROR
-
If not
recoverable perform Battery Low/Shut Down
Calibration (Calibration Section 4.9.4). If
problem persists replace
U503.
STACK OVERFLOW ERROR
-
Should always be
recoverable.
If the error persists replace
U503
or
U505.
BAD POINTER TO EEPROM ROUTINES
-
If not recoverable
replace
U505
then
U503.
INVALID KEY CODE ERROR
-
If not recoverable replace
U505
then
U503.
INVALID KEY SOURCE ERROR
-
If not recoverable
replace
U505
then
U503.
INVALID QUEUE POINTER ERROR
-
If not recoverable
replace
U505
then
U503.
BATTERY VALUE OUT OF RANGE
-
If not recoverable
perform Battery Low/Shut Down Calibration
(Calibration Section 4.9.4).
If problem persists
replace
U503.
ERROR PERTAINS TO FACTORY CALIBRATION PORT
-
If not
recoverable replace
U503.
54c