TM 262
3-24 Revised 4/11/08
Printing test results
The printout will begin with a header if it is selected during the program mode (i.e., TM 262 or
a custom header designed by you). The next two lines contain space for entering the individuals
name and the test date. This is followed by the test results in the order that they were obtained/
selected.
Either a single test can be printed from memory or the entire group of tests in memory can be
printed. To print a single test from memory, use the PAGE button to arrive at the desired test
result to print. Once this test is displayed, press the PRINT SCREEN button.
To print all tests in memory, press the PRINT ALL button. When PRINT ALL is pressed and
two audiogram tests are stored in memory, they will combine under the following conditions.
There must be one left test and one right test sequentially stored in memory. A left and right
audiometric pair of tests will not be combined if they are separated in the memory by a Tymp
test. Therefore, when tests are erased, the resorting could cause a change in left, right or right,
left sequence with Audiometric tests. This would result in the wrong audiometric tests being
combined when PRINT ALL is selected. Prior to selecting PRINT ALL you should scroll
through the tests in memory to determine where the audiometric tests are located. This will help
you avoid combining tests from different patients.