EasyManuals Logo

Welch Allyn TM 262 User Manual

Welch Allyn TM 262
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
TM 262
5-6 Revised 4/11/08
31 2 sChar Slow Index The data index where the last compliance data point
measured at 600 daPa/sec before the rate changes to
200 daPa/sec is stored.
-1 = No rate change occurred
33 2 sChar Fast Index The data index where the first compliance data
point measured at 600 daPa/sec after changing back
to 600 daPa/sec from the 200 daPa/sec rate is
stored.
-1 = No rate change back occurred
35 2uCharTymp Data [0]
Tympanometry compliance data point #0 in
cm
3
,
stored as: compliance X 64. Range = 0.00 to 3.98
cm
3
.
A maximum of 88 data points are stored per
Tympanometry test.
:::: :
209 2 uChar Tymp Data [87]
Tympanometry compliance data point #87 in
cm
3
,
stored as: compliance X 64
211 2uCharTymp Scale
Tympanometry compliance axis scale. 15 = 1.5
cm
3
30 = 3.0
cm
3
213 2 uChar Number of
Reflex Tests
The number of reflex tests performed Range = 0 to
4
215 2 uChar Reflex Test
Parameters
Reflex test selection parameters.
Bit 0 = Ispi status:
0 = Not selected
1 = Selected
Bit 1 = Contra status:
0 = Not selected
1 = Selected
Bit 2 = 500 Hz status:
0 = Not selected
1 = Selected
Bit 3 = 1000 Hz status:
0 = Not selected
1 = Selected
Bit 4 = 2000 Hz status:
0 = Not selected
1 = Selected
Bit 5 = 4000 Hz status:
0 = Not selected
1 = Selected
Bits 6 & 7 = Not used
Character
Number
Number of
Characters
Data
Type Field Name Field Description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn TM 262 and is the answer not in the manual?

Welch Allyn TM 262 Specifications

General IconGeneral
BrandWelch Allyn
ModelTM 262
CategoryMedical Equipment
LanguageEnglish

Related product manuals