14
seriously.
Error 3
Remove any
clothing
interfering with
the cuff.
The signal is poor
during blood
pressure
measurement.
Take blood
pressure on a
bare arm.
Error x (x>4)
Contact
customer
service.
The device has
malfunctioned.
contact
customer
service.
Check connect
The ECG cable is
off, or the body
parts are not place
on the electrodes
closely.
Check the ECG
cable, or place
the electrodes
close to your
body part.
ECG waveform
drifts
The pressure
exerted on the
electrode is not
stable or too
much.
Hold the device
stably and
gently.
Hand or body may
be moving.
Try to keep
perfectly still
and test again.
The ECG
waveform
amplitude is
small
The measurement
method you
choose is not
suitable for you.
Change another
lead and try
again.
5. Accessories
Model
Description
540-00240-00
MICRO USB charge cable

Related product manuals