Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m d = 1.2 P
80 MHz to 800 MHz
d = 2.3 P
800 MHz to 2.5 GHz
Where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufactur-
er and d is the recommended
separation distance in metres
(m).