28
Electromagneticimmunity
TheRC-DEXisintendedforuseintheelectromagneticenviron-
ment specified below The customer or the user of an RC-DEX
shouldassurethatitisusedinsuchanenvironment
Immunity
Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Electro-
static
discharge
(ESD)
IEC
61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic
material, the
relative humidity
should be at
least 30 %.
Electrical
fast
transients/
burst
IEC
61000-4-4
± 2 kV for
power line
supplies
± 1 kV for
input/
output lines
Not
applicable *)
Not
applicable *)
Not
applicable *)
Surge
IEC
61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV
line(s) to
earth
Not
applicable *)
Not
applicable *)
Not
applicable *)