EasyManua.ls Logo

Widex RC-Dex - Voltage Variation Immunity; Power Frequency Magnetic Field Immunity

Widex RC-Dex
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
Voltage
dips, short
inter-
ruptions and
voltage
variations
on power
supply input
lines
IEC
61000-4-11
<5 % U
T
(>95 % dip
in U
T
) for
0.5 cycle
40 % U
T
(60 % dip
in
U
T
) for 5
cycles
70 % U
T
(30 % dip
in U
T
) for 25
cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 s
Not
applicable *)
Not
applicable *)
Power
frequency
(50/60 Hz)
magnetic
field
IEC
61000-4-8
3 A/m 3 A/m Power frequency
magnetic fields
should be at
levels
characteristic of
a typical location
in a typical
commercial or
hospital
environment
NOTE U
T
is the a.c. mains voltage prior to the application of
the test level.
*) Battery powered equipment

Other manuals for Widex RC-Dex