30
Electromagneticimmunity–cont
TheRC-DEXisintendedforuseintheelectromagneticenviron-
ment specified below The customer or the user of an RC-DEX
shouldassurethatitisusedinsuchanenvironment
Immunity
Test
IEC 60601
Test level
Compli-
ance
level
Electromagnetic en-
vironment – guid-
ance
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the RC-DEX,
including cables, than
the recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Recommended
separation distance
d = 1.2
P