EasyManua.ls Logo

Widex RC-Dex - Radiated RF Immunity and Separation

Widex RC-Dex
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Immunity
Test
IEC 60601
Test level
Compli-
ance
level
Electromagnetic en-
vironment – guid-
ance
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m d = 1.2
P
80 MHz to 800 MHz
d = 2.3
P
800 MHz to 2.5 GHz
Where P is the
maximum output
power rating of the
transmitter in watts
(W) according to the
transmitter
manufacturer and d is
the recommended
separation distance in
metres (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey
a
, should be less
than the compliance
level in each frequency
P range
b
.
Interference may
occur in the vicinity of
equipment marked
with the following
symbol: