Electromagneticimmunity
TheTV-DEXisintendedforuseintheelectromagneticenvironmentspecified
belowThecustomerortheuserofaTV-DEXshouldassurethatitisusedin
suchanenvironment
Immunity Test IEC 60601
Test level
Compliance level Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30 %.