46 WIKA operating instructions, model CPG1200
EN
14603921.02 06/2023 EN/DE
Calculation factor of the user-defined unit
To change the calculation factor of the user-defined unit, proceed as follows under
“Menu” / “Settings” / “User unit”:
1. With the [LIGHT/▼] button or [ZERO/▲] button, select the factor to be changed.
2. Confirm the selection with the [MENU/◀▶] button.
⇒
An input field for the calculation factor appears. (The input is made from left to
right.)
3. Use the [LIGHT/▼] button or [ZERO/▲] button to select the number and use the
[MENU/◀▶] button to jump one digit to the right.
The input can be aborted with the [Home] button. The instrument exits the
menu.
To accept the set calculation factor, press the [MENU/◀▶] button repeatedly until the
instrument jumps back one menu level.
Press the [HOME] button in order to exit the menu.
There are 7 significant digits available for entering the conversion factor,
plus decimal point, plus sign (+/-).
The conversion is always made for the unit “bar”, independent from the set indication
unit.
The conversion is done using the following formula:
Current measured value (bar) x conversion factor
Example:
Current measured value in “bar” = 5.123 bar
Conversion factor user unit1: +0.264
Measured value in “User unit1” = 5.113 * 0.264 = 1.352
Designation of the user-defined unit
To change the designation of the user-defined unit, proceed as follows under “Menu” /
“Settings” / “User unit”:
1. With the [LIGHT/▼] button or [ZERO/▲] button, select the unit to be changed.
2. Confirm the selection with the [MENU/◀▶] button.
⇒
An input field for the limit value appears. (The input is made from left to right.)
3. Use the [LIGHT/▼] button or [ZERO/▲] button to select the letter and use the
[MENU/◀▶] button to jump one digit to the right.
7. Operation via menu functions