52 WIKA operating instructions, model CPG1200
EN
14603921.02 06/2023 EN/DE
7.7 Firmware update
A firmware update can be made via the free software packages, WIKA-Cal or
WIKA-DCS.
Procedure
1. Install the WIKA-Cal software or alternative WIKA-DCS on a PC or notebook.
2. Connect the instrument to the PC or notebook via the USB interface.
3. Open the software
4. Open window X
5. Upload the firmware to the instrument
Do not disconnect the USB connection while the firmware is being transferred to
the instrument.
6. Switch off the instrument
7. Switch on the instrument again
Due to the firmware update, the switch-on process is delayed by approx. 3
seconds.
7. Operation via menu functions / 8. Faults
8. Faults
Personnel: Skilled personnel
Protective equipment: Protective gloves, safety goggles
Tools: SW 27 open-ended spanner or torque wrench
WARNING!
Physical injuries and damage to property and the environment
caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
Should a failure occur, hazardous media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
Use the requisite protective equipment.