53WIKA operating instructions, model CPG1200
EN
14603921.02 06/2023 EN/DE
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
10.2 “Return”.
For contact details, see chapter 1 “General information” or the back page of
the operating instructions.
Display Causes Measures
Low battery voltage, functioning is
only guaranteed for a short period
of time
Insert new batteries, see chapter
9.2 “Battery replacement”
OL Measuring range far exceeded
≥ 5 % FS
The pressure source must be
removed immediately from the
CPG1200 to prevent damage to the
internal sensor.
OL
-OL
Reading is significantly above or
below the measuring range ≥ 5 % FS
Check: Is the pressure within the
permissible measuring range of the
sensor?
No display or
instrument is
not responding
to button press
Battery is empty Insert new batteries, see chapter
9.2 “Battery replacement”
Batteries inserted incorrectly Ensure the correct polarity, see
chapter 9.2 “Battery replacement”.
System error Switch off the CPG1200, wait for a
short period of time, switch on again
Defect in the CPG1200 Send in for repair
8. Faults