26 WIKA operating instructions digital pressure gauge, model DG-10
11250313.04 01/2023 EN/DE/FR/ES
EN
8. Dismounting, return and disposal / 9. Specications
Do not dispose of with household waste. Ensure a proper disposal in accordance with
national regulations.
9. Specifications
Specications
Measuring range → See product label
Digital indicator → See chapter 5.3 “Description of the digital indicator”
Software functions → See chapter 5.4 “Key function in the operating mode”
Voltage supply → See chapter 5.2 “Voltage supply”
Reference conditions
Temperature 0 ... 60 °C [32 ... 140 °F]
Air pressure 860 ... 1,060 mbar
Air humidity < 90 % rel., non-condensing
Nominal position Calibrated in vertical mounting position with process connection facing
downwards.
Supply voltage 2 x 1.5 V AA cell
Accuracy
Accuracy at reference
conditions
≤ ±0.5 % of span ±1 digit
Including non-linearity, hysteresis, zero oset and end value deviation
(corresponds to measured error per IEC 61298-2).
Calibration accuracy of zero
signal
≤ 0.1 % of span (power-up reset)
Adjustability of zero point ≤ 20 % of span (via tare function with model DG-10-E)