Pressure gauge with reed contact model 851-A, for Ex areas
EN
Safety instructions for installation
Install instruments in accordance with the manufacturer's
instructions and the valid standards and regulations.
Only connect circuits with the same voltage and type of
protection to the switch contacts and connection leads of the
instrument.
Limit the maximum current, using external measures, to a
value of ≤ 100 mA per circuit.
Size the connection leads for the largest current strength in the
circuits and ensure sufficient UV resistance and mechanical
stability.
With flexible connection leads, use isolated end splices. Max.
permissible conductor cross-section 1.5 mm².
Connection leads must be suited to the ambient temperature
range of the application.
The connection leads must also be suitable for the supplied
cable gland.
Seal the cable entry with the appropriate approved cable
glands.
Install the connection cables securely.
11449056.01
Cable outlet