wtw pH 340/ION - User Manual

wtw pH 340/ION
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...

Table of Contents

Questions and Answers

  • B
    Brian HillAug 25, 2025
    What to do if wtw Measuring Instruments display 0000FFFFLLLL Indication range exceeded?
    • J
      Jennifer WoodardAug 25, 2025
      If your wtw Measuring Instruments display '0000FFFFLLLL Indication range exceeded', it suggests a problem with the pH electrode. The issue might be that the electrode is not immersed, the cable is broken, the electrolytic gel has dried out, there's an air bubble in front of the diaphragm, or there's air in the diaphragm. To solve this: * Immerse the electrode into the sample solution. * If needed, exchange the electrode. * Remove any air bubbles. * Suck off air or moisten the diaphragm.
  • A
    ashleypalmerNov 9, 2025
    What to do if wtw pH 340/ION show EEEE3333 error?
    • T
      Tammy GarnerNov 10, 2025
      If your wtw Measuring Instruments display the error 'EEEE3333' indicating that the slope or asymmetry is outside the acceptable ranges, it could be due to several reasons: * The pH electrode might be too old, the diaphragm or membrane could be contaminated, there might be moisture in the plug, or the electrolytic gel could be too old. * The meter may have an incorrect calibration procedure, the solution temperature might be wrong, or there could be moisture in the socket. * The buffer solution could be unsuitable, too old, or used up. To resolve this: * Exchange the pH electrode, clean the diaphragm and membrane, and dry the plug. If needed, exchange the electrode. * Select the correct calibration procedure, set the correct temperature, and dry the socket. * Use WTW technica...

wtw pH 340/ION Specifications

General IconGeneral
Brandwtw
ModelpH 340/ION
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals