4. OPERATION PROCEDURE
Monitoring Data
: Heater Output
Heater output shows heater power output ratio in a
percentage of rated capacity. Output is controlled
by a PID operation value between 100 and 0%
until temperature setting is reached.
Example
If upper display shows , ouput is at 45.6% of rated
heater capacity.
: Accumulated power-on time (in hours).
First digit of 5 shown (0xxxx)
Power-on time is elapsed time between turning
main power swtich (ELB) ON () and OFF ().
: Accumulated power-on time (in hours).
Last four digits of 5 shown (x0000)
Example
If accumulated power-on time is 50,067 hours the first
screen will show 5 and next screen
will show 67. Maximum total for this indicator is 65,535.
: Accumulated operation run time (in hours).
First digit of 5 shown (0xxxx)
Accumulated operation run time is the sum total
of hours aggregated between the start and end of
operation runs.
: Accumulated operation run time (in hours).
Last four digits of 5 shown (x0000)
Example
If accumulated operation run time is 10,023 hours, the
first screen will show 1 and next
screen will show 23. Maximum total for this indicator is
65,535.