10. SPECIFICATIONS
Specifications Table (continued from previous page)
Model Name
DF412
DF612
DH412
DH612
Control
Devices &
Functions
Additional functions
Variable Air Flow Function
Power-on Time and Operation Time Accumulation Monitor (up to 65,535
hours); Calibration Offset; Monitoring Display for Accumulated Power
Consumption, Total CO2 Emissions, and Heater Operation Output; Power
Recovery Mode; Setting Data Backup and Recovery
Heater Control
Triac with Zero-cross Control
Sensor
K type Thermocouple double sensor
(for temperature control and independent overheat prevention device)
Safety Device
Control Board
Self-diagnosis Functions (Detection for Temp. Sensor Failure, TRIAC Short
Circuit, Heater Line Disconnect, Fan Motor Failure, Main Relay Contact
Damage and Overheating), Key Lock Function
Earth Leakage
Breaker (ELB)
15A
20A
20A
30A
Leak Current/Short Circuit/Over-current Protection,
Rated Current Sensitivity 30mA
Independent
Overheat
Prevention Device
(IOPD)
Temperature Setting Range:
0300°C
Temperature Setting Range:
0400°C
Door switch
Door open: fan motor and heater circuit OFF
Door close: fan motor and heater circuit ON
Standard
Weights & Measurements
Interior
Dimensions Width
Depth
Height
*3
450mm
450mm
450mm
600mm
600mm
600mm
450mm
450mm
450mm
600mm
600mm
600mm
Overall
Dimensions Width
Depth
Height
*3
1050mm
630mm
850mm
1200mm
780mm
1000mm
1050mm
630mm
850mm
1200mm
780mm
1000mm
Internal Capacity
91
216
91
216
Weight
Approx. 112kg
Approx.156kg
Approx.112kg
Approx.156kg
Number of tiers/
rack support pitch
9 tiers/45mm
9 tiers /60mm
9 tiers /45mm
9 tiers /60mm
Chamber Rack
Load Capacity
Approx. 30kg
/ rack
Approx.30kg
/rack
Approx.30kg
/rack
Approx.30kg
/rack
Included
Items
Chamber Racks /
Rack Supports
2/4
3/6
2/4
3/6
Instruction Manual
1 copy
Notes
*1 Performance based on rated source voltage, single phase 220V±5%, room temperature 23°C
±5°C, 65RH±20 humidity, 86kPa to 106kPa atmospheric pressure, exhaust damper
closed, no sample load.
*2 Value calculated by dividing JIS measurement by 2.
*3 Protrusions excluded.