XK3190-A12(E)
The AD value is too small when calibrating the full capacity
Please change proper capacity load cell
The zero point is out of range when calibrating the zero point
Please make sure no load is on the scale
The zero point is out of range when turning on
Please make sure no load is on the scale when turning on
The input sample quantity is zero when sampling in counting mode
Please input the right sample quantity
When full capacity calibrating, the weight input is zero
Please input the weight in accordance with the load on scale
The unit weight is less than 0.25e when sampling in counting mode
Please re-input the sample quantity
The weight is beyond the semi-automatic zero-setting range, the [Zero] key is
invalid.
Overflow of the accumulated weight
Please clear the accumulated weight
The G.W. is less than -20e
Overload, or load cell malfunction
Please recalibrate the F.S., and check the connection of load cell