XK3190A12(E)
16
ERR 1
The AD value is too small when calibrating the full capacity
Please change proper capacity load cell
ERR 2
The zero point is out of range when calibrating the zero point
Please make sure no load is on the scale
ERR 3
The zero point is out of range when turning on
Please make sure no load is on the scale when turning on
ERR 4
The input sample quantity is zero when sampling in counting mode
Please input the right sample quantity
ERR 5
When full capacity calibrating, the weight input is zero
Please input the weight in accordance with the load on scale
ERR 6
The unit weight is less than 0.25e when sampling in counting mode
Please re-input the sample quantity
ERR 7
The weight is beyond the semi-automatic zero-setting range, the [Zero] key is
invalid.
ERR 8
Overflow of the accumulated weight
Please clear the accumulated weight
bAt-lo
Low power
Lo
The G.W. is less than -20e
oL
Overload, or load cell malfunction
Please recalibrate the F.S., and check the connection of load cell