11-10
IM 04L42B01-01E
• ErrorsRelatedtoFTPClient
For a description of the FTP client function of the DX, see the Communication Interface
User’s Manual, IM 04L41B01-17E. The detail code does not appear in the error message
on the screen. You can view the code on the FTP log display of the DX or using the FTP
log output via communications.
Code Message
280 IP address is not set or FTP function is not available.
Further details are provided by the character string that appears after error code 280.
CharacterStringandDetails
HOSTADDR
An IP address has not been assigned to the DX.
Check the IP address.
DORMANT
Internal processing error.
*1
LINK
Data link is disconnected.
Check the cable connection.
281 FTP mail box operation error.
Further details are provided by the character string that appears after error code 281.
CharacterStringandDetails
MAIL
Internal processing error.
*1
STATUS
Internal processing error.
*1
TIMEOUT
Internal processing error.
*1
PRIORITY
Internal processing error.
*1
NVRAM
Internal processing error.
*1
282 FTP control connection error.
Further details are provided by the character string that appears after error code 282.
CharacterStringandDetails
HOSTNAME
Failed the DNS lookup (search the IP address corresponding to the host name).
Check the DNS setting and the destination host name.
TCPIP
Internal processing error.
*1
UNREACH
Failed to connect to a control connection server.
Check the address setting and that the server is running.
OOBINLINE
Internal processing error.
*1
NAME
Internal processing error.
*1
CTRL
The control connection does not exist.
Check that the server does not drop the connection and that it responds within the proper time
period.
IAC
Failed to respond in the TELNET sequence.
Check that the server does not drop the connection and that it responds within the proper time
period.
ECHO
Failed to transmit data on the control connection.
Check that the server does not drop the connection and that it responds within the proper time
period.
11.1AListofMessages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Daqstation DX2020 and is the answer not in the manual?

YOKOGAWA Daqstation DX2020 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDaqstation DX2020
CategoryRecording Equipment
LanguageEnglish

Related product manuals