Calibrating after Adjustment
After adjustment, be sure to calibrate the instrument. After calibration, enter the user calibration date.
Entering the User Calibration Date
1.
Press MENU. The top menu appears.
2.
Press the Utility soft key, the Next soft key, and then the Next soft key. The following menu appears.
3.
Press the Maintenance soft key. The Maintenance menu appears.
4.
Press the Calibration Date soft key. The following menu appears.
5.
If you calibrated the D/A output, press the D/A Cal. Date soft key. The following screen appears.
6.
Use the arrow(▲,▼,◄,and►)keystoenterthedate,andthenpressENTER.
7.
If you calibrated the voltage/current measurement (/DM option), in step 5 above, press the DMM Cal.
Date soft key, and enter the calibration date.