<10. Other Functions>
10-30
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
Item Initialization Parameter Default setting
Alarm setting
(continued on next
page)
Hysteresis O2 concentration hysteresis 0.1%O
2
Humidity hysteresis 0.1%H
2
O
Mixing ratio hysteresis 0.001kg/kg
Relative humidity hysteresis 0.1%R.H.
Alarm operation delay
Alarm operation delay 3s
Oxygen
concentration alarm
Oxygen
concentration alarm
(HH) high-high alarm 󰀨
(HH) high-high alarm value
100.0%O
2
(H) high alarm 󰀨
(H) high alarm value 100.0%O
2
(L) low alarm 󰀨
(L) low alarm value 0.0%O
2
(LL) low-low alarm 󰀨
(LL) low-low alarm value 0.0%O
2
Humidity alarm Humidity alarm (HH) high-high alarm 󰀨
(HH) high-high alarm value
100.0%H
2
O
(H) high alarm 󰀨
(H) high alarm value 100.0%H
2
O
(L) low alarm 󰀨
(L) low alarm value 0.0%H
2
O
(LL) low-low alarm 󰀨
(LL) low-low alarm value 0.0%H
2
O