Technical parametersCommon fault and trouble shooting
Warning: If the situations cannot be solved
or unexpected problem happens, please
consult the local distributor.
Fault cause
Wrong
indication
Solution
Err 4
Err 5
Err 6
Err 7
Err 8
Showing low
voltage icon
" "
Showing the
" "icon
Unable to measure
pressure
Pressurizing error
Fasten cuff correctly
before measurement
Check if there is air
leakage from the cuff
Pressurizing error
caused by arm or
body motion
Keep arm and body still
and measure again
Cuff is too lose or
fall off
Fasten cuff tightly
Pressure exceeds the
maximum value
(290mmHg)
Battery is low
Battery is running out
Measure again please
Replace new batteries
Replace new batteries
The table below shows the possible fault
displaying icon, possible reason, and solutions.
36
37
YE670D
LCD digital display
Oscillation mensuration
0-280mmHg
0-450mmHg
40-200 time/min
Pressure:±3mmHg(±0.4kPa)
Pulse:±5% of reading value
=Type BF application part
4X1.5v AA batteries
300 times
22~32cm
IP21
Approx.267g
Approx.127x93x74 (mm)
+10℃ to +40℃/15% to 90%RH
(non-condensing)
-20℃ to +55℃/15% to 90%RH
(non-condensing)
5 years or 50 thousand times
Model
Display
Measuring
method
Measuring range
Cuff Pressure
Pulse rate
Precision
Electric
classification
Power supply
Battery life
Cuff size
IP Classification
Service life
Weight
Dimension
Operating
temperature
/Humidity
Storage
temperature
/Humidity
YE670D( )