Addendum Table of contents
Instructions for use and disinfection of
tonometer measuring prisms
319300-9231-000-AddGA01-GB-240320 3 / 18
Table of contents
1 Notes on the instructions for use ............................................................... 5
1.1 Scope of application .......................................................................................................... 5
1.2 Intended use of the document........................................................................................... 5
1.3 Questions and comments .................................................................................................. 5
1.4 Conventions in this document ........................................................................................... 5
1.4.1 Conventions in all text areas.......................................................................................................... 5
1.4.2 Conventions in a course of action.................................................................................................. 6
2 Safety notes ............................................................................................... 7
2.1 Ambient conditions ........................................................................................................... 7
2.2 Important general advice................................................................................................... 7
2.3 Operation, environment .................................................................................................... 7
2.4 Warranty and product liability........................................................................................... 7
2.5 Validated cleaning and disinfecting agents ....................................................................... 8
2.6 List of disinfectants with good material compatibility with ZEISS measuring prisms......... 9
2.7 Service life ......................................................................................................................... 12
3 Cleaning and disinfection of reusable measuring prisms............................ 13
3.1 Manual cleaning steps ....................................................................................................... 13
3.2 Disinfection steps .............................................................................................................. 14
3.3 Rinsing and storage........................................................................................................... 15

Table of Contents